5520A-SC300 Option
Verification Tables
8
8-89
8-115. Voltage Function Verification: DC Voltage into a 1 M
Load
Nominal Value
(dc) Measured Value (dc) Deviation (mV)
1-Year Spec.
(mV)
0.0 mV 0.10
5.0 mV 0.11
-5.0 mV 0.11
22.0 mV 0.15
-22.0 mV 0.15
25.0 mV 0.16
-25.0 mV 0.16
45.0 mV 0.21
-45.0 mV 0.21
50.0 mV 0.23
-50.0 mV 0.23
220.0 mV 0.65
-220.0 mV 0.65
250.0 mV 0.72
-250.0 mV 0.72
450.0 mV 1.22
-450.0 mV 1.22
500.0 mV 1.35
-500.0 mV 1.35
3.3 V 8.35
-3.3 V 8.35
4.0 V 10.10
-4.0 V 10.10
33.0 V 82.60
-33.0 V 82.60