Loading...
Introduction and Specifications
Additional Specifications
1
1-33
AC Current (Non-Sine Wave) Specifications (cont)
Square Wave
Range
p-p Frequency
1-Year Absolute
Uncertainty,
tcal ± 5 °C,
± (% of output + % of range)
Max
Current
Resolution
0.01 Hz to 10 Hz 5.0 + 0.5 Two digits
0.047 mA to 10 Hz to 45 Hz 0.25 + 0.5
0.65999 mA [1] 45 Hz to 1 kHz 0.25 + 0.25 Six digits
1 kHz to 10 kHz 10 + 2
0.01 Hz to 10 Hz 5.0 + 0.5 Two digits
0.66 mA to 10 Hz to 45 Hz 0.25 +0.5
6.59999 mA [1] 45 Hz to 1 kHz 0.25 + 0.25 Six digits
1 kHz to 10 kHz 10 + 2
0.01 Hz to 10 Hz 5.0 + 0.5 Two digits
6.6 mA to 10 Hz to 45 Hz 0.25 + 0.5
65.9999 mA [1] 45 Hz to 1 kHz 0.25 + 0.25 Six digits
1 kHz to 10 kHz 10 + 2
0.01 Hz to 10 Hz 5.0 + 0.5 Two digits
66 mA to 10 Hz to 45 Hz 0.25 + 0.5
659.999 mA [1] 45 Hz to 1 kHz 0.25 + 0.5
1 kHz to 10 kHz 10 + 2
0.66 A to 10 Hz to 45 Hz 0.5 + 1.0 Six digits
5.99999 A [2] 45 Hz to 1 kHz 0.5 + 0.5
1 kHz to 10 kHz 10 + 2
6 A to 41 A [2] 45 Hz to 500 Hz 0.5 + 0.5
500 Hz to 1 kHz 1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.