Fluke 5520A - 6-74. SC600 Hardware Adjustments; 6-76. Adjusting the Leveled Sine Wave Function

Fluke 5520A
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Operation
Checking 5520A Status
5
5-43
15 14 13 12 11 10 9 8
0 0 RPTBUSY SETTLED REMOTE 0 UUTBFUL UUTDATA
76543210
HIVOLT MAGCHG TMPCAL 0000
OPER
Set to 1 when a calibration report is being printed to the serial port.
Set to 1 when the output has stabilized to within speclfication or the TC measurement
has settled and is available.
Set to 1 when the 5520A is under remote control.
Set to 1 when data from the UUT port has filled up the UUT buffer.
Set to 1 when there ia data available from the UUT port.
Set to 1 when the 5520A is programmed to a voltage above 33 Volts.
Set to 1 when the output magnitude has changed as a result of another change
(e.g. RTD_TYPE). This bit is always 0 in the ISR. It changes to 1 only in the
ISCR0 and ISCR1 registers.
Set to 1 when the 5520A is using temporary (non-stored) calibration data.
Set to 1 when the 5520A is in operate, 0 when it is in standby.
RPTBUSY
SETTLED
REMOTE
UUTBFUL
UUTDATA
HIVOLT
MAGCHG
TMPCAL
OPER
nn320f.eps
Figure 5-11. Bit Assignments for the ISR, ISCEs and ISCRs