5520A-SC300 Option
Summary of Commands and Queries
8
8-85
SCOPE?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Returns the oscilloscope’s current mode of operation. Returns OFF if the oscilloscope is
off.
Parameter: (None)
Response: <character> (Returns OFF, VOLT, EDGE, LEVSINE, MARKER, or
WAVEGEN.)
TRIG
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
Programs the oscilloscope’s trigger output BNC.
Parameters: OFF (Turns the trigger output off.)
DIV1 (Turns the trigger output on. Frequency is the same as the
signal at SCOPE output.)
DIV10 (Turns the trigger output on. Frequency is 1/10 of the
signal at SCOPE output.)
DIV100 (Turns the trigger output on. Frequency is 1/100 of the
signal at SCOPE output.)
TRIG?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Returns the output setting of the oscilloscope’s trigger.
Parameters: (None)
Response: <character> (Returns OFF, DIV1, DIV10, or DIV100.)
OUT_IMP
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Programs the oscilloscope’s output impedance.
Parameters: Z50 (Programs the oscilloscope’s output impedance to
50.)
Z1M (Programs the oscilloscope’s output impedance to
1 M.)
OUT_IMP?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
Returns the impedance setting of the oscilloscope’s output.
Parameters: (None)
Response: <character> (Returns Z50 or Z1M.)