5520A
Operators Manual
4-30
13. The Control Display now shows your entries. For example, 123.456 mV and
234.567 mA at 1.1234 kHz (below).
nn086f.eps
14. Press E. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
nn087f.eps
15. Press O to activate the calibrator output. When changing power output levels,
you must reenter both voltage and current (in either order).
(Enter voltage or current and then a watts entry value using b A. The remaining
volts or current value is calculated and displayed.)
Three softkey labels appear on the Control Display: WAVE MENUS, I OUT (AUX or
20A terminals), and LCOMP (off or on). The Control Display also shows the real power
output for sine waves. Power out is computed as Power = Cosine Φ (Volts x Current)
where Φ is the phase difference between the volts and current waveforms. Cosine Φ is
also known as the Power Factor (PF).