Front Panel Operation
Setting the Output
4
4-39
7. The Control Display now shows the amplitude of your capacitance entry. For
example, 123.45 µF (below).
nn099f.eps
8. Press E. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
nn100f.eps
9. Press O to activate the calibrator output.
The softkey in the Control Display labeled COMP allows you to select one of three lead-
compensation settings.
nn101f.eps
COMP (Compensation) Applies 2-wire compensation or turns compensation off.
Compensation refers to methods of connecting the 5520A to the UUT to cancel out
test lead resistance (NOT capacitance). Compensation is available for capacitances
of 110 nF and above. This softkey will not function below 110 nF. See “Four-Wire
versus Two-Wire Connections” earlier in this chapter for more information.