Front Panel Operation
Setting the Output
4
4-41
Note
The entered temperature will be cleared to 0
°
C (32
°
F) if you change
between tc and rtd, or change the type of thermocouple (except for a type B
thermocouple, which clears to 600
°
C). If this should occur, select
OUTPUT tc, the desired thermocouple TYPE, and then reenter the
temperature.
nn104f.eps
Out@TC terminal (Output at the front panel TC terminals) Displays the actual dc
voltage at the front panel TC terminals. This is a display only, not a softkey function.
TC MENUS (Thermocouple Menu) Shows submenus for thermocouple outputs.
UNITS (Temperature Units) Selects °C or °F as the temperature unit.
REF SRC (Reference Source) Selects intrnl (Internal) or extrnl (External)
temperature reference source. Select intrnl when the selected thermocouple has