Front Panel Operation
Setting the Output
4
4-43
6. The Control Display now shows the amplitude of your temperature output. For
example, 123.456 °C (below).
nn102f.eps
7. Press E. The calibrator clears your entry from the Control Display and copies
it into the Output Display (below is typical).
nn103f.eps
8. Press O to activate the calibrator output.
Four softkey labels appear on the Control Display. Press the OUTPUT softkey to toggle
the rtd selection, displaying the rtd setup menu and four softkey positions.
Note
The temperature you entered above will be cleared to 0
°
C (32
°
F) if you
change between tc (thermocouple) and rtd (resistance temperature
detector), or change the type of rtd. If this occurs, select OUTPUT rtd, the
desired rtd TYPE, and then reenter the temperature following steps 4 to 8.
nn105f.eps