Front Panel Operation
Setting the Output
4
4-45
4-34. Measuring Thermocouple Temperatures
Complete the following procedure to measure the output of a thermocouple connected to
the TC input. If you make an entry error, press G to clear the display, then reenter.
1. Press R to clear any output from the 5520A.
2. Connect the thermocouple to the front panel TC connector.
Note
Use thermocouple wire and miniconnectors that match the type of
thermocouple. For example, type K wire and type K miniconnectors.
3. Press U to display the TC menus (below).
nn106f.eps