5520A
Operators Manual
5-14
nn308f.bmp
5. Select the Communications command from the Setting menu. Enter the RS-232
parameters that match those selected at the Calibrator for the Host port. If using the
5520A factory defaults, the Communications dialog box for COM1 will appear as
shown below. Select COM as required. Click OK.
nn309f.bmp
6. Verify the Calibrator is powered and in the reset condition. (If in doubt, press R
on the Calibrator front panel.)
7. On the Terminal screen, type the command REMOTE and press <Enter>. Observe
the Calibrator Control Display changes to REMOTE CONTROL (below).
nn325f.eps
The characters REMOTE should have appeared on the terminal screen as they were
entered. If they did not appear on the screen, but the Control Display changed to
REMOTE CONTROL, then refer to step 4 of the “RS-232 Host Port Setup
Procedure” and change the REMOTE I/F setting from comp to term
.