Output Commands (cont)
DPF Sets the displacement power factor (phase angle) between the NORMAL and AUX
terminals for ac power output only.
DPF? Returns the displacement power factor (phase angle) between the NORMAL and
AUX terminals.
DUTY Sets the duty cycle of square wave outputs.
DUTY? Returns the duty cycle of square wave outputs.
FUNC? Returns the present output, measurement, or calibration function.
HARMONIC Makes the frequency of one output be a harmonic (multiple) of the other output
(called the fundamental).
HARMONIC? Returns the present instrument harmonic and fundamental locations.
LCOMP Activates or deactivates inductive load compensation for ac current output.
LCOMP? Returns whether inductive load compensation for ac current output is active.
OPER Activates the Calibrator output if it is in standby.
OPER? Returns the operate/standby setting.
OUT Sets the output of the Calibrator and establishes a new reference point for the error
mode.
OUT? Returns the output amplitudes and frequency of the Calibrator.
PHASE Sets the phase difference between the NORMAL and AUX terminals for dual
outputs. The NORMAL terminal output is the phase reference.
PHASE? Returns the phase difference between the NORMAL and AUX terminals.
POWER? Returns the equivalent power for dc and ac power output.
RANGE? Returns the present output ranges.
RANGELCK Locks in the present range, or selects auto ranging.
RANGELCK? Returns whether or not the preset output range is locked.
REFCLOCK Sets the reference clock source (internal or through the 10 MHz IN BNC
connector).
REFCLOCK? Returns the reference clock source (internal or through the 10 MHz IN BNC
connector).
REFPHASE If two Calibrators are synchronized using 10 MHz IN/OUT, sets the phase
difference between the NORMAL terminals on the slave Calibrator and the
NORMAL terminals of the master Calibrator.
REFPHASE? If two Calibrators are synchronized using 10 MHz IN/OUT, returns the phase
difference between the NORMAL terminals on the slave Calibrator and the
NORMAL terminals of the master Calibrator.
STBY Puts the Calibrator in standby.
SYNCOUT Sends a synchronization pulse out to a slave Calibrator through the 10 MHZ OUT
BNC connector.
WAVE Sets the waveforms for ac outputs.
WAVE? Returns the waveforms of the output.
ZCOMP Activates (2-wire or 4-wire) or deactivates impedance compensation.
ZCOMP? Returns whether or not impedance compensation is active and if active, which
type.