Remote Commands
Commands
6
6-13
DBMZ_D
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(dBm Impedance Default command) Sets the power-up and reset default impedance
used for dBm outputs (ac volts).
Parameters: Z50 (50 ohms)
Z75 (75 ohms)
Z90 (90 ohms)
Z100 (100 ohms)
Z135 (135 ohms)
Z150 (150 ohms)
Z300 (300 ohms)
Z600 (600 ohms)
Z900 (900 ohms)
Z1000 (1000 ohms = dBv)
Z1200 (1200 ohms)
Example: DBMZ_D Z600
This setting only applies when single output AC voltages are being sourced. The dBm
impedance is set to the default at power on, reset, and when going into single output AC
mode.
DBMZ_D?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(dBm Impedance Default query) Returns the power-up and reset default impedance used
for dBm outputs (ac volts).
Response: (character) Impedance keyword
Example: DBMZ_D? returns Z600
DC_OFFSET
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(DC Voltage Offset command) Applies a dc offset to an ac output voltage (maximum
six digits). This command applies only to single ac voltage outputs. If the selected offset
is too large for the active ac voltage range, an error message is returned.
Parameter: <value> signed offset amplitude
Example: DC_OFFSET +123.45 MV
Load a dc offset of +123.45 mV to the ac output signal.
DC_OFFSET?
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(DC Voltage Offset query) Returns the value of the dc offset voltage.
Response: <value> signed offset amplitude