Remote Commands
Commands
6
6-27
Example: MULT 2.5
Multiply the existing reference by 2.5, creating a new reference. For example, an
existing reference of 1 V is multiplied to 2.5 V.
NEWREF
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(New Reference command) Sets the new reference to the present Calibrator output value
and exit the error mode (if selected). For example, you might edit the Calibrator output
using the EDIT and INCR commands, and then use the NEWREF command to establish a
new reference point and exit the error mode. This is the same as pressing the Calibrator
front panel N key.
Parameter: (None)
Example: NEWREF
Set the reference value to the current Calibrator output value.
OLDREF
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Old Reference command) Sets the Calibrator output to the reference value and exit the
error mode (if selected). If editing the output using the EDIT and INCR commands and
you want to return to the reference value, use the OLDREF command. If editing the
output and you want to make the edited value the new reference, use the NEWREF
command.
Parameter: (None)
Example: OLDREF
Set the output to the existing reference value, clearing editing changes.
ONTIME?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Calibrator On Time query) Returns the time in minutes since the Calibrator was most
recently powered up.
Response: <days>,<hours> (24-hour clock)
Example: ONTIME? returns 47
Returns the time since the Calibrator was last powered up: 47 minutes.
*OPC
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Operations Complete command) Sets bit 0 (OPC) of the Event Status Register to 1
when all pending device operations are complete. Also see the *ESR? command.
Parameter: (None)
Example: *OPC
Set bit 0 of the Event Status Register to 1 when all pending device operations are done.