5520A
Operators Manual
6-40
RPT_STR
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Report String command) Loads the user report string. The user report string can be read
on the Control Display in local operation, and appears on calibration reports. The
CALIBRATION switch must be set to ENABLE. (Sequential command.)
Parameter: String of up to 40 characters
RPT_STR?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Report String query) Returns the user report string. The user report string can be read
on the Control Display in local operation, and appears on calibration reports. (Sequential
command.)
Parameter: None
Response: (String) Up to 40 characters
*RST
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xxx
(Reset Instrument command) Resets the Calibrator to the power-up state. *RST holds
off execution of subsequent commands until the reset operation is complete. This
command is the same as pressing the front panel R key.
A reset action evokes the following commands and values:
Command Value Command Value
CUR_POST AUX REFCLOCK <REFCLOCK_D value>
DBMZ <DBMZ_D value> REFPHASE <REFPHASE_D value>
DC_OFFSET 0V RTD_TYPE <RTD_TYPE_D value>
DUTY 50PCT SCOPE OFF
EARTH OPEN STBY (No output)
EXTGUARD OFF TC_OFFSET 0 CEL
HARMONIC 1, PRI TC_OTCD ON
LCOMP OFF TC_REF INT
LOWS TIED TC_TYPE <TC_TYPE_D value>
OUT 0V,0HZ TRIG OFF
OUT_IMP Z1M TSENS_TYPE TC
PHASE 0DEG WAVE NONE,NONE
PRES_UNIT <PRES_UNIT_D value> ZCOMP OFF
RANGELCK OFF ZERO_MEAS OFF
Changes made to the setup menus that are not saved in memory are discarded on reset.
Response: (None)