5520A
Operators Manual
6-42
Parameters: PT385 (100-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_200 (200-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_500 (500-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_1000(1000-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT3926 (100-ohm RTD, curve α=0.003926 ohms/ohm/°C)
PT3916 (100-ohm RTD, curve α=0.003916 ohms/ohm/°C)
CU10 (10-ohm RTD, empirical curve)
NI120 (120-ohm RTD, empirical curve)
Example: RTD_TYPE_D PT3926
Set the RTD default type to a 100-ohm RTD with curve α=0.003926 ohms/ohm/°C.
RTD_TYPE_D?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Resistance Temperature Detector Type Default query) Returns the default Resistance
Temperature Detector (RTD) used at power on and reset.
Responses: PT385 (100-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_200 (200-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_500 (500-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT385_1000(1000-ohm RTD, curve α=0.00385 ohms/ohm/°C)
PT3926 (100-ohm RTD, curve α=0.003926 ohms/ohm/°C)
PT3916 (100-ohm RTD, curve α=0.003916 ohms/ohm/°C)
CU10 (10-ohm RTD, empirical curve)
NI120 (120-ohm RTD, empirical curve)
Example: RTD_TYPE_D? returns PT3926
Returns PT3926 when the RTD default type is a 100-ohm RTD with curve
α=0.003926 ohms/ohm/°C.
SP_SET
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Host Serial Port Set command) Sets the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port, which is saved in the Calibrator non-volatile
memory. (While saving configuration data in the non-volatile memory, a period of about
2 seconds, the Calibrator does not respond to remote commands.) (To set the parameters
for the rear panel SERIAL 2 TO UUT serial port, see the UUT_SET command.) The
factory default values are shown below in bold type. (To return to the factory defaults,
see the FORMAT SETUP command.)
The interface selection sets the command response, with command echo back for
commands and error messages with TERM (terminal) or no echo back with COMP
(computer).
Parameters: <baud rate value>, 300, 600, 1200, 2400, 4800, 9600
<interface>, TERM (terminal), COMP (computer)
<flow control>, XON (xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>, DBIT7 (7 bits) or DBIT8 (8 bits)
<number stop bits>, SBIT1 (1 bit) or SBIT2 (2 bits)
<parity>, PNONE (none), PODD (odd),PEVEN (even)
<end of line char.> CR (carriage return), LF (line feed),
CRLF (carriage return/line feed)