5520A
Operators Manual
6-56
*WAI command is useful with any overlapped command, preventing the Calibrator from
processing other commands until the overlapped command is processed.
Example: *WAI
Process all existing commands before continuing.
WAVE
Overlapped CoupledIEEE-488 RS-232 Sequential
xx xx x
(Waveform command) Sets the waveforms for ac outputs. If the Calibrator is sourcing
one output, one parameter is required. If the Calibrator is sourcing two outputs, two
parameters are required or one parameter to set the waveform to both outputs. Waveform
choices are SINE (sine wave), TRI (triangle wave), SQUARE (square wave), TRUNCS
(truncated sine wave), or NONE (waveform does not apply).
Parameter: <1st waveform> , (SINE, TRI, SQUARE, TRUNCS, NONE)
<2nd waveform> (SINE, TRI, SQUARE, TRUNCS, NONE)
Example: WAVE SINE,SQUARE
Set the waveforms for a dual output to Sine wave on the primary output (Calibrator front
panel NORMAL terminals) and Square wave on the secondary output (front panel AUX
or 20A terminals).
WAVE?
Sequential Overlapped
Coupled
IEEE-488 RS-232
x x xx
x
(Waveform query) Returns the waveform types for ac outputs. Waveform choices are
SINE (sine wave), TRI (triangle wave), SQUARE (square wave), TRUNCS (truncated
sine wave), or NONE (waveform does not apply).
Responses: <1st waveform> , (SINE, TRI, SQUARE, TRUNCS, NONE)
<2nd waveform> (SINE, TRI, SQUARE, TRUNCS, NONE)