Table 7-3. Verification Tests for DC Voltage (AUX)
Range Output Lower Limit Upper Limit
329.999 mV 0.000 mV -0.350 mV 0.350 mV
329.999 mV 329.000 mV 328.551 mV 329.449 mV
329.999 mV -329.000 mV -329.449 mV -328.551 mV
3.29999 V 0.33000 V 0.32955 V 0.33045 V
3.29999 V 3.29000 V 3.28866 V 3.29134 V
3.29999 V -3.29000 V -3.29134 V -3.28866 V
7.0000 V 7.0000 V 6.9976 V 7.0025 V
7.0000 V -7.0000 V -7.0025 V -6.9976 V