Maintenance
Performance Tests
7
7-11
Table 7-5. Verification Tests for Resistance (cont)
Range Output Lower Limit Upper Limit
10.99999 M 3.30000 M 3.29959 M 3.30041 M
10.99999 M 10.90000 M 10.89875 M 10.90125 M
32.99999 M 11.90000 M 11.89512 M 11.90488 M
32.99999 M 19.00000 M 18.99370 M 19.00630 M
32.99999 M 30.00000 M 29.99150 M 30.00850 M
109.9999 M 33.0000 M 32.9838 M 33.0162 M
109.9999 M 109.0000 M 108.9534 M 109.0466 M
329.9999 M 119.0000 M 118.6025 M 119.3975 M
329.9999 M 290.0000 M 289.1750 M 290.8250 M
1100.000 M 400.000 M 394.700 M 405.300 M
1100.000 M 640.000 M 631.820 M 648.180 M
1100.000 M 1090.000 M 1076.420 M 1103.580 M