EasyManua.ls Logo

Fluke 5520A - Page 275

Fluke 5520A
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance
Performance Tests
7
7-13
Table 7-6. Verification Tests for AC Voltage (Normal) (cont)
Range Output Frequency Lower Limit Upper Limit
3.29999 V 3.00000 V 450 kHz 2.99340 V 3.00660 V
3.29999 V 3.29000 V 2 MHz 0.07500 V (Note)
32.9999 V 3.3000 V 45 Hz 3.2990 V 3.3010 V
32.9999 V 3.3000 V 10 kHz 3.2990 V 3.3010 V
32.9999 V 30.0000 V 9.5 Hz 28.3350 V 31.6650 V
32.9999 V 30.0000 V 10 Hz 29.9919 V 30.0082 V
32.9999 V 30.0000 V 45 Hz 29.9957 V 30.0044 V
32.9999 V 30.0000 V 1 kHz 29.9957 V 30.0044 V
32.9999 V 30.0000 V 10 kHz 29.9957 V 30.0044 V
32.9999 V 30.0000 V 20 kHz 29.9928 V 30.0072 V
32.9999 V 30.0000 V 50 kHz 29.9904 V 30.0096 V
32.9999 V 30.0000 V 90 kHz 29.9759 V 30.0241 V
329.999 V 33.000 V 45 Hz 32.993 V 33.007 V
329.999 V 33.000 V 10 kHz 32.989 V 33.011 V
329.999 V 300.000 V 45 Hz 299.953 V 300.047 V
329.999 V 300.000 V 1 kHz 299.953 V 300.047 V
329.999 V 300.000 V 10 kHz 299.946 V 300.054 V
329.999 V 300.000 V 18 kHz 299.928 V 300.072 V
329.999 V 300.000 V 50 kHz 299.922 V 300.078 V
329.999 V 200.000 V 100 kHz 199.630 V 200.370 V
1020.00 V 330.00 V 45 Hz 329.91 V 330.09 V
1020.00 V 330.00 V 10 kHz 329.91 V 330.09 V
1020.00 V 1000.00 V 45 Hz 999.74 V 1000.26 V
1020.00 V 1000.00 V 1 kHz 999.79 V 1000.21 V
1020.00 V 1000.00 V 5 kHz 999.79 V 1000.21 V
1020.00 V 1000.00 V 8 kHz 999.74 V 1000.26 V
1020.00 V 1020.00 V 1 kHz 1019.79 V 1020.21 V
1020.00 V 1020.00 V 8 kHz 1019.74 V 1020.27 V
Note: Typical specification is -24 dB at 2 MHz

Table of Contents

Other manuals for Fluke 5520A

Related product manuals