8-3. Volt Specifications
Table 8-1. Volt Specifications
Volt Function
dc Signal Square Wave Signal [1]
50 Load 1 M Load 50 Load 1 M Load
Amplitude Characteristics
Range
0 V to ±6.6 V 0 V to ±130 V ±1 mV to
±6.6 V p-p
±1 mV to
±130 V p-p
Resolution
Range
1 mV to 24.999 mV
25 mV to 109.99 mV
110 mV to 2.1999 V
2.2 V to 10.999 V
11 V to 130 V
Resolution
1 µV
10 µV
100 µV
1 mV
10 mV
Adjustment Range Continuously adjustable
1-Year Absolute Uncertainty,
tcal ± 5 °C
± (0.25% of
output +
40 µV)
± (0.05% of
output +
40 µV)
± (0.25% of
output + 40
µV)
± (0.1% of
output +
40 µV) [2]
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Characteristics
Range 10 Hz to 10 kHz
1-Year Absolute Uncertainty,
tcal ± 5 °C
± (2.5 ppm of setting)
Typical Aberration
within 4 µs from 50% of
leading/trailing edge
< (0.5% of output + 100 µV)
[1] Selectable positive or negative, zero referenced square wave.
[2] For square wave frequencies above 1 kHz, ± (0.25% of output + 40 µV).