EasyManua.ls Logo

Fluke 5520A - Page 293

Fluke 5520A
420 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5520A-SC600 Option
SC600 Option Specifications
8
8-9
8-6. Time Marker Specifications
Table 8-4. Time Marker Specifications
Time Marker into
50 5 s to 50 ms
20 ms to
100 ns
50 ns to
20 ns 10 ns
5 ns to
2 ns
1-Year Absolute
Uncertainty at
Cardinal Points,
tcal ± 5 °C
± (25 + t
*1000) ppm [1]
± 2.5 ppm ± 2.5 ppm ± 2.5 ppm ± 2.5 ppm
Wave Shape spike or
square
spike, square,
or 20%-pulse
spike or
square
square or
sine
sine
Typical Output
Level
> 1 V p-p [2] > 1 V p-p [2] > 1 V p-p [2] >1 V p-p [2] > 1 V p-p
Typical Jitter (rms) <10 ppm < 1 ppm < 1 ppm <1 ppm <1 ppm
Sequence 5-2-1 from 5 s to 2 ns (e.g., 500 ms, 200 ms, 100 ms )
Adjustment
Range [3]
At least ± 10% around each sequence value indicated above.
Amplitude
Resolution
4 digits
[1] t is the time in seconds.
[2] Typical rise time of square wave and 20%-pulse (20% duty cycle pulse) is < 1.5 ns.
[3] Time marker uncertainty is ± 50 ppm away from the cardinal points.
8-7. Wave Generator Specifications
Table 8-5. Wave Generator Specifications
Wave Generator Characteristics
Square Wave, Sine Wave, and Triangle Wave
into 50 or 1 M
Amplitude
Range into 1 M: 1.8 mV to 55 V p-p
into 50 : 1.8 mV to 2.5 V p-p
1-Year Absolute Uncertainty,
tcal ± 5 °C, 10 Hz to 10 kHz
± (3% of p-p output + 100 µV)
Sequence 1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Typical DC Offset Range 0 to ± (40% of p-p amplitude) [1]
Frequency
Range 10 Hz to 100 kHz
Resolution 4 or 5 digits depending upon frequency
1-Year Absolute Uncertainty, tcal ± 5 °C ± (25 ppm + 15 mHz)
[1]The DC offset plus the wave signal must not exceed 30 V rms.

Table of Contents

Other manuals for Fluke 5520A

Related product manuals