5520A
Operators Manual
8-22
Output @ SCOPE
terminal (50Ω).
MORE
OPTIONS
SET TO
LAST F
MODE
levsine
fffff
(see
“The
MORE
OPTIONS
Menu”)
LAST F
50 kHz
volt
edge
levsine
marker
wavegen
video
pulse
meas Z
overld
You can press the MODE softkey to cycle through the functions in the order shown, or
you can press P to return directly to the SCOPE menu.
Each option in the LEVSINE menu is described below:
• OUTPUT @ SCOPE terminal (50Ω) Indicates the location and impedance of the
signal output. If the signal does not appear on the oscilloscope, press O. To
disconnect the signal, press Y. You cannot change the impedance while you are
in LEVSINE mode.
• MORE OPTIONS Opens additional menu items, which are described in detail
under “The MORE OPTIONS Menu.”
• SET TO LAST F Toggles between the current frequency setting and the reference
value of 50 kHz. This option is useful for reverting to the reference to check the
output after you make adjustments at another frequency.
• MODE Indicates you are in LEVSINE mode. Use the softkey to change modes and
open menus for other calibration modes.
8-35. Shortcuts for Setting the Frequency and Voltage
The following three options are available for controlling the sine wave settings:
• SET TO LAST F toggles between the last frequency used and the reference
frequency of 50 kHz, letting you check the output at the reference after you make
adjustments at a different frequency.
• MORE OPTIONS lets you use an automatic frequency sweep and lock the voltage
range, if necessary. The following section provides details on this menu.
• The X and D keys step frequencies up or down in amounts that let you
quickly access a new set of frequencies. For example, if the value is 250 kHz, X
changes it to 300 kHz, and D changes it to 200 kHz. For voltage values, X
and D step through cardinal point values in a 1.2-3-6 sequence.