5520A-SC600 Option
Remote Commands and Queries
8
8-37
TRIG?
(IEEE-488, RS-232, Sequential)
Returns the output setting of the oscilloscope’s trigger.
Parameters: (None)
Response: <character> (Returns OFF, DIV1, DIV10, or DIV100.)
OUT_IMP
(IEEE-488, RS-232, Sequential)
Programs the oscilloscope’s output impedance.
Parameters: Z50 (Programs the oscilloscope’s output impedance to
50.)
Z1M (Programs the oscilloscope’s output impedance to
1 M.)
Example: OUT_IMP Z50
OUT_IMP?
(IEEE-488, RS-232, Sequential)
Returns the impedance setting of the oscilloscope’s output.
Parameters: (None)
RANGE
(IEEE-488, RS-232, Sequential)
Programs the instrument range in PULSE, MEAS Z, OVERLD modes.
Parameters:
Pulse TP0DB TP8DB TP20DB TP28DB TP40DB TP48DB
Range 2.5 V 1.0 V 250 mV 100 mV 25 mV 10 mV
Impedance
Measure
TZ50OHM TZ1MOHM TZCAP
Range res 50 res 1M cap
Overload TOLDC TOLAC
Range DC AC
Example: RANGE TP20DB