5520A
Operators Manual
3-22
Table 3-3. Factory Defaults for SETUP Menus Power-Up Defaults
Parameter Setting
SETUP Menu in
Figure 3-4.
User report string (*PUD
string)
Cleared.
D
Error units > 0.1%
G1
SC-600 option overload
test safety timeout
10 s
G1
Temperature standard its-90
G1
Host interface gpib (IEEE-488)
G1
UUT serial interface 8 bits, 1 stop bit, xon/xoff, parity
none, 9600 baud
I
Host serial interface term, 8 bits, 1 stop bit, xon/xoff,
parity none, 9600 baud, CRLF,
012,000
K, L, M, N
GPIB Port Address 4
O
Display brightness (Note) level 1,0
P
Display contrast (Note) level 7,7
P
dBm impedance 600
S
Pressure units psi
S
RTD type pt385
S1
Thermocouple type K
S1
Phase reference 0.00°
S3
10 MHz reference clock internal
S2
Current limits ±20.5 A
U
Voltage limits ±1020 V
V
Note: Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6,7.