EasyManua.ls Logo

Fluke 6080A - A1 Display PCA 6-3.; A2 Coarse Loop PCA 6-4.; A3 Sub-Synthesizer VCO PCA 6-5.; A4 Sub-Synthesizer PCA 6-6.

Fluke 6080A
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TROUBLESHOOTING AND REPAIR
A1 Display PCA 6-3.
Adjustments: None.
A2 Coarse Loop PCA 6-4.
Adjustments: None.
Perform Reference Oscillator Calibration. See paragraph 3-14.
A3 Sub-Synthesizer VCO PCA 6-5.
Adjustments: R106 on Sub-Synthesizer PCA. See paragraph 6C-9.
A compensation data EPROM containing VCO tuning data is included. See
paragraph 6C-21 for data transfer instructions.
A4 Sub-Synthesizer PCA 6-6.
Adjustments: R106. See paragraph 6C-9.
Perform Reference Oscillator Calibration. See paragraph 3-14.
A5 Coarse Loop VCO PCA 6-7.
Adjustments: None.
Perform Coarse Loop compensation. See Appendix H.
A6 Mod Oscillator PCA 6-8.
Adjustments: None.
A7 Relay Driver PCA 6-9.
Adjustments: None.
A8 Output PCA 6-10.
Adjustments:
R28, detector offset/linearity. See paragraph 6D-13 (Mod Control PCA).
R20, RF Level adjustment. See paragraph 6D-15 (Mod Control PCA).
RIO, AM Depth adjustment. See paragraph 6D-14 (Mod Control PCA).
R96, Q16 Bias adjustment. See paragraph 6D-21 (Output PCA).
A compensation data EPROM containing Output PCA level correction data is
included. See paragraph 6-21 for data transfer instructions.
A9 Sum Loop VCO PCA 6-11.
Adjustments: R51 and C7, AM Bandwidth adjust. See paragraph 6D-20 (Premodulator
PCA).
Perform Sum Loop compensation. See Appendix H.
6-5

Table of Contents

Related product manuals