EasyManua.ls Logo

Fluke 6080A - STATUS SIGNALS AND STATUS CODES 6-35.

Fluke 6080A
296 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TROUBLESHOOTING AND REPAIR
Table 6-13. Filter Tests
STATUS SIGNALS AND STATUS CODES 6-35.
Table 6-14. lists the major hardware status signals monitored by the software and the
corresponding front panel status code.
Table 6-14. Status Signals and Codes
6-13
CODE
339
340
341
342
343
344
345
346
347
348
349
350
351
352
353
354
355
356
FREQUENCY
20 MHz
30 MHz
40 MHz
60 MHz
100 MHz
150 MHz
200 MHz
300 MHz
400 MHz
550 MHz
650 MHz
800 MHz
100 MHz
100 MHz
100 MHz
200 MHz
500 MHz
1024
MHz
FREQ BAND
15 - 22 MHz
22 - 32 MHz
32 - 47 MHz
47 - 64 MHz
64 - 128 MHz
128 - 180 MHz
180 - 256 MHz
256 - 350 MHz
350 - 512 MHz
512 - 625 MHz
625 - 730 MHz
730 -
1056
MHz
15 - 22 MHz
22 - 32 MHz
32 - 47 MHz
47 - 64 MHz
256 - 350 MHz
512 - 730 MHz
EXPECTED STATE
OF ALC LOOP
Leveled
Leveled
Leveled
Leveled
Leveled
Leveled
Leveled
Leveled
Leveled
Leveled
Leveled
Leveled
Unleveled
Unleveled
Unleveled
Unleveled
Unleveled
Unleveled
Amplitude = 15.0 dBm
ASSEMBLY
A7 Attenuator/RPP
A10 Premodulator
A4 Sub-Synthesizer
A2 Coarse Loop
A12 Sum Loop
A12 Sum Loop
A2 Coarse Loop
A14 FM loop
STATUS
CODE
240
241
242
243
244
245
246
247
SIGNAL
RPTRPL
ALCUNLVL
SUBUNLKL
CORUNLKL
SUMUNLKL
SUMUNLVL
REFUNLKL
FMUNLKL
DESCRIPTION
RPP Tripped
ALC Loop Unleveled or AM
Overmodulation
Sub-Synthesizer Unlocked
Coarse Loop Unlocked
Sum Loop Unlocked
Sum Loop Unleveled
Reference Loop Unlocked
FM Loop Unlocked or FM
Overmodulation

Table of Contents