PERFORMANCE TESTS
Table 4-1. Recommended Test Equipment
INSTRUMENT NAME
DVM
DMM
RMS Voltmeter
Wideband Amplifier
RF-Spectrum Analyzer
Oscilloscope
FET Probe
500-ohm Probe
RF Voltmeter
Frequency Counter
Modulation Analyzer
Distortion Analyzer
Power Meter
Power Sensor
(High-level)
Power Sensor
(Low-Level)
Attenuator, 50, 20 dB
LF Synthesized Signal
Generator
HF Synthesized
Sig-Gen
(Low Residual)
MINIMUM REQUIREMENT
5 1/2-Digit, 0.3% DC-20 kHz
3 1/2-Digit, 1% DC and 1 kHz
10 Hz to 20 MHz, low noise
> 25-dB gain, 0.4 to 1050 MHz
NF < 9dB
0.1 to 1.7
GHz,
100 Hz BW
Four-trace 300 MHz, 5-mV/Div
DC-900 MHz
DC-3.5 GHz, 10X
0.01
to 700
MHz, 0.01
to 3V ± 10%
0.1-1050 MHz; 10 Hz res; 0.1V
Input: 0.15 to 1300 MHz, 0 to +20 dBm
AM: 10 to
90%, ±1%,
FM: 0.1 to 100 kHz dev ±1%
External LO capability
1 to 10%
rng,
±1 dB, 0.4 and 1 kHz
Instrumentation accuracy < ±1%
-30 to 20 dBm; VSWR < 1.2 for
0.4 to 1
MHz,
< 1.1 for 1 to
2000 MHz, < 1.3 for > 2000 MHz
-67 to -20 dBm; VSWR < 1.4 for
10 to 30 MHz <
1.15
for 30 to
2100
MHz
0.1 to 2100 MHz VSWR < 1.15
10 Hz to 11 MHz, 10 Hz steps,
1V peak, Spurs and Harm < -50 dB
0.5 to
1024
MHz
MANUFACTURER
DESIGNATION
JF 8840A-09
JF 8020B
JF 8922A
HP 8447D-010
HP 8568A
TEK 2465-11
TEK
6201
TEK P6156
HI RF 801
JF 7220A
HP 8901A
w/Option -003
HP 339B
HP 436A
HP 8482A
HP 8484A
Narda 777C
JF
6011A
JF 6080A/AN
NOTES
1
A,P
A,P,T
P
P,T
T,P
T
T
T
2
A,P,T
A,P,T
A,P,T
A.P.T
P
3
A,P
P,T
4-2

Table of Contents