ACCESS PROCEDURES
WARNING
PIVOTING MODULE INSTRUCTIONS.
THE SYNTHESIZER MODULE, WHICH MUST BE RAISED TO GAIN
ACCESS TO MANY OF THE CIRCUIT BOARDS, IS HEAVY. WHEN RAISING
OR LOWERING THE MODULE, OBSERVE THE FOLLOWING PROCE-
DURES TO AVOID INJURY:
RAISING THE MODULE:
1. REMOVE THE THREE #8 PAN-HEAD SCREWS THAT SECURE THE
MODULE TO THE CHASSIS SIDES.
2. GRASP THE HANDLE AND LIFT THE MODULE.
3. LOCK THE MODULE IN THE UP POSITION BY INSERTING TWO OF THE
PREVIOUSLY REMOVED #8 SCREWS INTO THE BOSSES PROTRUDING FROM
THE CHASSIS SIDES NEAR THE HINGES.
LOWERING THE MODULE:
1. SUPPORT THE MODULE IN THE RAISED POSITION AND REMOVE THE TWO
LOCK-UP SCREWS.
2. USING THE HANDLE ONLY AND KEEPING HANDS CLEAR OF ALL OTHER
PARTS OF THE SIGNAL GENERATOR, LOWER THE MODULE.
3. LOCK THE MODULE IN THE DOWN POSITION, USING THE THREE #8 PAN-HEAD
SCREWS.
CAUTION
The gas spring can make the synthesizer module swing open when the
instrument is turned on its side. To avoid this, be certain to lock the
synthesizer module in the down position after lowering it.