101364     Model FT2 
4.1.5.2.  For Discrete Output Setting: 
 
Press ALM (F2) key to select alarm output, then menu will display: 
 
 
 
ALM = HiFloAlm 
NXT           OK 
F1  F2  F3  F4 
 
 
 
 
 
 
Press NXT (F1) repeatedly until the correct selection is shown and then press OK (F4) to 
accept the setting. 
 
Selections are:  Not used 
HiFloAlm  for High Flow Alarm 
LoFloAlm  for Low Flow Alarm 
HiTempAlm  for High Temperature Alarm 
LoTempAlm  for Low Temperature Alarm 
 
      If a Limit type is selected, then the menu will display: 
 
 
 
 
 
 
 
HiFloAlm = 1000 SCFM 
CHG                      OK 
F1  F2  F3  F4 
 
Enter the limit value to trigger the contact alarm output and then press OK (F4) to accept 
the setting. 
 
Press EXIT (F4) repeatedly until “Normal Mode” is seen briefly to exit the programming 
mode. 
 
The JP1 selection jumper needs to be installed between pin 2 & 3 on the main FT2        
board to select the isolated digital output for the alarm. 
 
4.1.6.  RS485 Serial Communication Settings 
 
    To program the RS485 settings, press I/O (F1) key from the base menu 
 
           SET  I/O 
I/O      FEQ       420         EXIT 
F1  F2  F3  F4 
 
 
 
 
 
 
 
 Then press I/O (F1) again: 
Fox Thermal Instruments Inc., 399 Reservation Road, Marina, CA 93933  Page 33