EasyManuals Logo

Fox FT2 User Manual

Default Icon
56 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
101364 Model FT2
4.1.8.1. To Program the Password:
Press PSW (F2) key to select password.
This screen displays the current Level 1 password.
Press CHG (F1) key to change the password and enter new value accordingly to 4.1.1
description of entering data.
Press OK (F4) to accept new data and exit programming by pressing EXIT (F4) key
repeatedly until out of the programming mode.
Note: Password can be number or letter characters up to 4 digits.
PASSWD = 1234
CHG OK
F1 F2 F3 F4
4.1.9. Units Settings
This menu is used to set the units for mass flow, temperature, pressure reference and the settings
of reference temperature, reference pressure and density of gas when using Lbs/time or Kg/time.
These values will be set at FOX, using the Application Data Sheet values. If the customer changes
the application, these values can be changed to match the new application. Check with FOX
customer service before changing the application gas.
The unit setting is accessed from the base programming menu by pressing
FLO (F2) key:
SET PARAMETERS
I/O FLO DSP EXIT
F1 F2 F3 F4
FLOW PARAMETERS 1
SIM UNT PRM EXIT
F1 F2 F3 F4
Press UNT (F2) for Unit selection:
FLO = SCFM
NXT OK
F1 F2 F3 F4
Press NXT (F1) key to change selection and OK (F4) to accept it
Fox Thermal Instruments Inc., 399 Reservation Road, Marina, CA 93933 Page 37

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fox FT2 and is the answer not in the manual?

Fox FT2 Specifications

General IconGeneral
BrandFox
ModelFT2
CategoryMeasuring Instruments
LanguageEnglish