10.4 Setting of range (setting screen for input/output range)
Set the measuring unit, range, output mode and error handling for analog input/output.
10.4.1 Setting the input range: When setting the range for the input current
or input voltage.
Setting range: 0.000 to ±9999999999
0 100%
Input
Input
conversion
Full scale
20mA or 5V
Base scale
4mA or 1V
Input conversion =
Input (%) / 100 × (Full scale – Base scale) + Base scale
(1) Press the or key on the RANGE screen to
select “INPUT RANGE”.
Press the
ENT
key to display the INPUT RANGE
screen.
(2) Select “CH1” or “CH2” by the
or key and
press the
ENT
key.
(3) The cursor moves to “BASE SCALE”.
Press the
ENT
key, and you are ready to set the base
scale.
Move the digit by the
or key and enter numeric
values by pressing the
or key.
After entry, press the
ENT
key to set the base scale.