66 INF-TN1FSC-E
10.1.2 ZERO ADJUSTMENT: when performing zero adjustment
[Operation]
(1) Select “ZERO ADJUSTMENT” by the
or key
and press the
ENT
key. The zero adjustment screen
will appear.
(2) Select ZERO ADJUSTMENT, and press the
ENT
key.
Zero adjustment to be speci ed is carried out.
• [SET ZERO]
Perform zero adjustment in situation where the ow
is stopped.
The measurement indication should be at zero when
the
ENT
key is pressed.
This zero calibration operation should be performed
after stopping ow.
• [CLEAR]
Adjustment is cleared.
On this screen, zero point is set or cleared.
When PROCESS SETTING or measurement
method (page 93) is changed, perform zero
adjustment.
CAUTION