INF-TN1FSC-E 37
7.13 Completion of PROCESS SETTING
After the settings are completed, press the key
“DECISION” is reversed from white to blue.
Pressing the
ENT
key to complete settings, and then
returns to the “SITE SETUP” screen.
After mounting the sensor, perform zero point
calibration.
ENT
MENU MENU
Note) When the inner mounting diameter is 13mm, the sensor mounting method is 0.0mm or less de-
pending on the pipe materials.
[Unit: mm]
Necessary pipe thickness for uid water
CARBON STELL 2.15 FRP 3.21
STAINLESS STEEL 1.87 DUCTILE IRON 2.15
PVC 3.69 PEEK 3.69
COPPER 3.82 PVDF 3.69
CAST IRON 2.98 ACRYLIC 2.90
ALUMINUM 1.99 PP 3.69
When the sensor mounting length is 0.0mm or less, error of the measurement is approximately ±2
to 5%.