DescriptionStage
After the test has finished, the test results can be printed. The results printout
shows all registered flow kit bar codes, installation procedures, notes, and
test results. If page 8 does include any warning or other text the test failed.
If all flow cell tests passed the system can be used.
After:
If any part of the test fails, there is a troubleshooting guide available (see
Section 8.3 Troubleshooting a failed component test, on page 117).
Failure:
Prepare a test
Wash liquid and test solution are required for a component test. A Test solution prepara-
tion protocol template is available on the system CD. The volumes depend on the flow
kit used:
ÄKTA ready High Flow Kit 25 liters of wash liquid and 10 liters of test solution.
ÄKTA ready Low Flow Kit 10 liters of wash liquid and 5 liters of test solution.
Test solution for the High Flow Kit is prepared as follows: 10 liters of 1.0 M NaCl with 1%
acetone in purified water (i.e., 584 g NaCl + 100 ml acetone in 9.90 liters purified water).
For the Low Flow Kit, half the amounts are used.
The test solution should be freshly made, so the acetone does not evaporate.
Note:
ActionStep
Remove the protective end cap from
inlet 6 (arrow), and connect the wash
liquid.
Note:
Containers must not be placed at a
level lower than 1 m below the inlets
(i.e., floor level).
1
ÄKTA ready Operating Instructions 28960345 AD 109
8 Component test