Enter installation information
After pressing Run in the installation wizard, you are directed to the UNICORN start
protocol with panels containing data about the installation.
Start Notes
In the first panel, Start Notes, is a text area where notes about the installation can be
entered. Any notes written here will be included in the report.
Questions
The Questions panel has fields for entering product number and name, and identification
data for the different parts of the flow kit and the pH electrode (if used), There is also a
field for the test solution protocol number.
Much of the information can be conveniently scanned in using the supplied bar code
scanner. It can also be manually entered. The information supplied will be included in
the report.
Result name
The panel Result Name contains fields for the name of, and path to, the result file, where
results will be stored. Enter a proper name for the result file, in other case it will get the
name Wizard generated and you may end up with a lot of such files with the same
name.
The Result Name panel also has a Start button. When finished entering data, click Start
to start the test.
ÄKTA ready Operating Instructions 28960345 AD 111
8 Component test
8.1 Run a component test