Manual calibration of
UV stray light
When the stray light calibration is performed a warning message is displayed. The
warning message can be disabled.
Follow the instruction below to perform a UV stray light calibration.
ActionStep
Click Alarms&Mon and then StraylightCalib_UV_AT_131 in the Instructions
box.
1
•
Enabled is the default option. If you click Execute the stray light calibra-
tion will be performed and a warning message is displayed:
"Stray light reduction calibration. Do not perform any other instruction
during the next 15 seconds."
Click Acknowledge when 15 seconds have passed.
•
If you click Disabled and after that Execute the stray light calibration
will be performed without the warning message on the screen and
without entry in the logbook.
148 ÄKTA ready Operating Instructions 28960345 AD
10 Creating a method
10.1 The Method Editor
10.1.3 Other Method Editor features