ActionStep
The report mode can also be changed only for the present run. After the run
the setting is restored to the previous one.
2
Click Alarms&Mon and then Fine_Tune_Fail_Report in the Manual menu.
•
If you click Enable and after that Execute the Fine_tuning_flowme-
ter_failed warning will be displayed if the system fails to perform a fine
tuning.
•
If you click Disable and after that Execute the Fine_tuning_flowme-
ter_failed warning will not be displayed on the screen and there will be
no entry in the logbook.
Totalizers
A totalizer monitors the accumulated volumes passing through an actual inlet or outlet.
There are six inlet totalizers and six outlet totalizers. For each totalizer, there is a watch
instruction allowing for easy programming of needed actions triggered by specific events.
To define an action based on the totalizer for, for instance, Inlet 1:
1
Enable the totalizer by inserting instruction Inlet1_Totalizer (under FlowPath) and
setting it to Enabled.
2
Insert watch instruction Watch_Inlet1_Totalizer (under Watch) and set values for
Test (e.g. Greater_Than), Value (volume in liters), and Action (e.g. END_BLOCK).
To activate an alarm for an enabled totalizer, insert the Alarm_XXX_Totalizer instruction
(under Alarms&Mon) corresponding to the inlet/outlet in question. (Note that alarms
automatically sets the system in pause mode.)
The allowed noise, or error, for a totalizer can be defined using the appropriate
WatchPar_XXX_Tot instruction (under Alarms&Mon).
A totalizer can be reset to 0 liters by inserting an XXX_Totalizer_Reset instruction (under
FlowPath).
150 ÄKTA ready Operating Instructions 28960345 AD
10 Creating a method
10.1 The Method Editor
10.1.3 Other Method Editor features