ActionStep
Flow meter cell:
Align the holes in the body of the
flow meter cell (FE-141) with the
two knobs on the cabinet (see 1
and 2 in figure) and push the cell
into place.
3 A
Clean the contact surfaces of
the transducer connectors
(bayonet couplings) and apply
a layer of Vaseline.
3 B
Note:
Make sure that enough Vaseline is applied to the transducer connectors. (Any
surplus Vaseline will be squeezed out and can be removed with a tissue.)
Connect the transducers to the connectors at the top and bottom of the
flow cell (see arrows in figure).
3 C
On the back of the flow meter cell is a window for the temperature sensor
(3 in figure). The window is automatically aligned with the temperature sensor
when the flow meter cell is mounted. No further action is necessary to pre-
pare for temperature measurements.
Note:
Make sure the flow meter cell is properly pushed onto its mounting knobs. A
soft click indicates correct mounting. Improper mounting of the flow meter
cell may result in erroneous temperature measurements and leakage.
92 ÄKTA ready Operating Instructions 28960345 AD
6 Flow kit installation
6.3 Mounting the flow kit