Mount the outlet tubing
ActionStep
XV-051 to XV-
056
1
2
Mount the outlet manifold by insert-
ing each outlet tube into the corre-
sponding outlet pinch valve (valves
XV-051 to XV-056). Do not yet remove
the end caps.
1
Close each valve safety lock by turn-
ing the black handles 180 degrees.
2
Snap the air vent tubing (that runs
from the top of the air trap) onto the
holder (1 in figure) on top of the sys-
tem cabinet, and remove the protec-
tive end cap (2) from the end of the
air vent tubing.
3
Note:
Remember to remove the protective end cap from the air vent tubing. The air
trap will not work properly if the cap is left in place.
Finishing flow kit installation
Make sure that the safety locks on all
pinch valves have been properly
closed. There are 19 safety locks in
total.
Make sure that the three pressure
flow cells are in place and properly
connected.
Set all pinch valves to run position by
turning the Installation switch to
RUN.
Check that the end cap on the air
vent tubing of the air trap has been
removed.
ÄKTA ready Operating Instructions 28960345 AD 95
6 Flow kit installation
6.3 Mounting the flow kit