Special Applications
Page 6-12
6.3.3.
DICOM Setup
The Workstation must be configured to access the DICOM network. Configuration identifies the server to
the network and defines the DICOM network nodes to the Workstation. The DICOM features include:
Worklist
DICOM Print
Storage and Commitment
Query & Retrieve
MPPS
Press the “Setup” key on the Workstation membrane keyboard.
Use this screen to access the Worklist, DICOM Printers, Storage and Commitment Server, Query &
Retrieve, MPPS Server and Network setup screen.
Network Setup
Click Network & DICOM button. The Network & DICOM setup screen shows on the right monitor.
Input the Local AE Title” and Station Name. The default AE Title is Brivo OEC. You can change it
according to the DICOM services requirement.