EasyManua.ls Logo

GE LOGIQ 5

GE LOGIQ 5
452 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-28 Section 8-4 - Keyboard Block
8-4-7 TGC Knob Set
Purpose: This is a description on how to remove and replace TGC Knobs.
8-4-7-1 Tools
Common pillips screwdrivers.
8-4-7-2 Needed Manpower
1persons, 5 minutes + travel
8-4-7-3 Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
8-4-7-4 Removal Procedure
1.) Remove eight (8) TGC knobs from the upper side of the Keyboard. Refer to Figure 8-32.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-4-7-5 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-32 Removing TGC Knobs
Table 8-14 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2300000, Rev 1+, Section 8-4-7. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-7-4
TGC Knob Set validation

Table of Contents