Chapter 8 Replacement Procedures 8-47
8-5-5 Top Cover
Purpose: This is a description on how to remove and replace the Top Cover.
8-5-5-1 Tools
• Common pillips screwdrivers
8-5-5-2 Needed Manpower
• 1 person, 20 minutes + travel
8-5-5-3 Preparations
• Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
8-5-5-4 Removal Procedure
1.) Remove the Left Cover. Refer to section 8-5-1 on page 39.
2.) Remove the Right Cover. Refer to section 8-5-2 on page 41.
3.) Remove the Rear Cover. Refer to section 8-5-3 on page 43.
4.) Unscrew two screws (1-2) from the top side of the Top cover. Refer to Figure 8-52.
Figure 8-52 Unscrew 2 screws from the Top cover
(1)
(2)