EasyManua.ls Logo

GE LOGIQ 5 - Front Cover

GE LOGIQ 5
452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
Chapter 8 Replacement Procedures 8-51
8-5-7 Front Cover
Purpose: This is a description on how to remove and replace the Front Cover.
8-5-7-1 Tools
Common pillips screwdrivers
8-5-7-2 Needed Manpower
1person, 20 minutes + travel
8-5-7-3 Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
8-5-7-4 Removal Procedure
1.) Remove the Left Cover. Refer to section 8-5-1 on page 39.
2.) Remove the Right Cover. Refer to section 8-5-2 on page 41.
3.) Remove the Front Base Cover. Refer to section 8-5-7 on page 51.
4.) Unscrew four screws (1-4). Refer to Figure 8-56.
5.) Remove the Front Cover.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-5-7-5 Mounting procedure
Install the new parts in the reverse order of removal.
Figure 8-56 Unscrew 2 screws on the left side
Table 8-26 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2300000, Rev 1+, Section 8-5-7. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-8-2
Cover Parts Function Validation
Section 10-5-5
Physical Inspection
(1)
(2)
Left Side
(3),(4) are on the right side

Table of Contents