8-4-12 OP Panel LCD Encoder Assy
Purpose: This is a description on how to remove and replace OP panel encoder Assy.
8-4-12-1 Tools
Common pillips screwdrivers
8-4-12-2 Needed Manpower
1person, 15 minutes + travel
8-4-12-3 Preparations
Shut Down the System and switch off the Main Breaker at the rear as described in section 4-3-2 on
page 3.
Keyboard Assy should be removed before proceed. For more information, refer to KeyBoard Assy
on section 8-4-1 on page 15.
8-4-12-4 Removal Procedure
1.) Pull out five LCD encoder knobs (2-6) and one LCD Brightness Encoder from the front side of the
keyboard. Refer to Figure 8-45.
2.) Turn the Keyboard Assy upside down.
Figure 8-45 Pulling out LCD encoder knobs