DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-30 Section 8-4 - Keyboard Block
8-4-8-4 Removal Procedure (cont’d)
3.) Unscrew eleven (11) screws (1-11) to remove the A/N keyboard prefix bracket. Refer to Figure 8-35.
4.) Remove the A/N keyboard prefix bracket
5.) Unscrew eight (8) screws (1-8) to remove the the A/N keyboard Assy. Refer to section Figure 8-36.
6.) Lift the A/N keyboard Assy Carefully.
7.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-4-8-5 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-35 Unscrew 11 screws
Figure 8-36 Unscrew 8 screws
Table 8-15 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2300000, Rev 1+, Section 8-4-8. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-7-5
A/N Key assy validation
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)
(9)
(10)
(11)
(1)
(2)
(3)
(4)
(5)
(6)
(7)
(8)