EasyManua.ls Logo

GE LOGIQ 5 - Page 291

GE LOGIQ 5
452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE MEDICAL SYSTEMS
D
IRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
Chapter 8 Replacement Procedures 8-95
8-7-3-4 Removal Procedure (cont’d)
5.) From the right view, disconnect two connectors (9-10).
6.) Unscrew four (4) screws (11-14) to remove the Backplane Assy.
7.) Pull the Backplane assy out carefully.
8.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-7-3-5 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-94 Disconnect 2 connectors from the right
WARNINGWARNING
The weight of the Backplane with PCBs inside is approximately 20 kg. Two person is
needed in the next step.
Table 8-45 Functional Tests
Service Manual
Section Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2300000, Rev 1+, Section 8-7-3. Equipment
passed all required tests and is ready for use. “
Section 4-3-2
Power Off / Shutdown
Section 4-9-8
Backplane Assy function check procedure
Section 4-3-4
System B/M-Mode Checks
Section 4-3-5
System CFM and PWD Checks
Section 10-5-2
Functional Checks (See Also Chapter 4)
Section 10-6
Using a Phantom
(9)
(10)
(11)
(12)
(13)
(14)

Table of Contents