DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-166 Section 8-10 - Software Loading Procedure
Section 8-10
Software Loading Procedure
8-10-1 General
This describes a full system software loading (Base Software Load Image + LOGIQ 5 Application)
procedure.
Use this instruction when:
- Performing the HDD with new one (Base Software Load Image + Application)
- Performing the system revision-up (Application)
Reference :
The below is the support table between Application Software version and Base System
Software Load Image part number
8-10-2 Parts Required
Base System Software Load Image CD
LOGIQ 5 R4.x.x Application Software
Blank CD-R ( for patient image backup)
Service Dongle
NOTE: Patient Image Data will be deleted when all the drives are ghosted. Back up them on the CD-R
or MOD before starting software loading
NOTE: Insert the service dongle into ths service port located at the rear panel before installing the Base
System Software Load image
8-10-3 TimeRequired
Approximately 2 hour
Table 8-56 Base Software Load Image CD Part Numbers
Application
Software Version
Base Software Load Image Part P/N
5134559-x (BEP2)
5132191-x (BEP3)
R4.X.X
Supported Supported