GE MEDICAL SYSTEMS
DIRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
8-182 Section 8-11 - Installing Base Software Load Image
8-11-1 Installing Base Software Load Image CD (cont’d)
13.)The following screen appears.
14.)If you choose ‘A’, you will be asked to press ‘Y’ to confirm your selection.
NOTE: For versions of software R3.x.x, your presets will be automatically translated (carried forward)
when the R3.0.0 software is loaded onto your system. Therefore, you DO NOT have to load
your R3.X.X after the upgrade to R4.X.X softwaree
Figure 8-220 Selection Screen
Figure 8-221 Confirm Screen
NOTICE
Select A when installing the new HDD (BEP replacement).
Select B when replacing C:\Partition only. It means all customer data are NOT deleted. The Ghost pro-
cedure is performed with data of the drives D, E, and F left.
Table 2-59 Drive Partition
Drive Partition Usage
C System Files and Application
D User Defined and Preset
E Patient Archive data
F Swap