GE MEDICAL SYSTEMS
D
IRECTION 2300000, REVISION 2 LOGIQ™5 SERVICE MANUAL
Chapter 10 Care & Maintenance 10-3
Table 10-2 Customer Care Schedule
Service at Indicated Time Daily Monthly Quarterly
Per Facilities
QA Program
Notes
Clean Probes •* * or before each use
Clean Probe Holders
•
Clean Air Filter •
more frequently depending on
your environment
Inspect AC Mains Cable • Mobile Unit Check Weekly
Inspect Cables and Connectors •
Clean Console •
Clean Monitor and Touch Panel •
Inspect Wheels, Casters, brakes and
Swivel Locks
• Mobile Unit Check Daily
Check Control Panel Movement
• Mobile Unit Check Daily
Console Leakage Current Checks •
also after corrective
maintenance
Peripheral Leakage Current Checks •
also after corrective
maintenance
Surface Probe Leakage Current Checks •
also after corrective
maintenance
Endocavity Probe Leakage
Current Checks
•
also after corrective
maintenance
Transesphongeal Probe Leakage
Current Checks
•
also after corrective
maintenance
Surgical Probe Leakage
Current Checks
•
also after corrective
maintenance
Measurement Accuracy Checks •
also after corrective
maintenance
Probe/Phantom Checks •
also after corrective
maintenance
Functional Checks •
also after corrective
maintenance