om 5184516-100 Rev. 5 6-10
ILLUSTRATION 6-7
PROCESS IMAGE SCREEN
6-2Image Subtraction
NOTE: If images were acquired subtracted, the images will automatically be shown
subtracted when you enter Review mode.
NOTE: The Subtraction feature is an Advanced Digital Package item option.
The Subtraction feature is used for post–acquisition image subtraction. Any image
can be declared a mask image, and all subsequent images in the series will be
subtracted from it.
1. Select the MASK button when the desired mask image is displayed.
2. The current frame # should appear in the FRAME# box (to the right of the
MASK button).
NOTE: User may select subtraction button and the first image of the sequence will be
used as the mask.
NOTE: Individual images and fluoro images may not be subtracted.
PROCESS IMAGE
PULL UP MENU
FOR TRAINING PURPOSES ONLY!